Masimo Radical Signal Extraction Manual del operador Pagina 40

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 39
5-8 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 6-16-1
6
troubleshooting
g
Troubleshooting
The following chart describes what to do if the Radical-7 Pulse CO-Oximeter system
does not operate properly or fails.
PROBLEM POSSIBLE CAUSE(S) RECOMMENDATION
UNIT DOES NOT POWER
ON.
One or both of the fuses
have blown.
Replace the fuses.
UNIT POWERS ON BUT
THE GRAPHIC DISPLAY
IS BLANK.
The viewing contrast is
not correct.
Use the Backlight/Contrast button
to adjust the viewing angle. If the
condition persists, the unit requires
service.
CONTINUOUS SPEAKER
TONE.
Internal failure.
Unit requires service. Press the
Alarm Silence button to silence
the alarm. If alarm continues to
sound, power down unit and remove
Handheld battery if necessary.
BUTTONS DON’T WORK
WHEN PRESSED
Internal failure. Unit requires service.
DEFECTIVE SENSOR
MESSAGE
Sensor or cable is broken.
Visually check the sensor LED if it is
flashing on and off.
If not, reconnect the cable and check
the LED again.
If the LED still fails to come on,
replace the sensor and/or cable.
The following chart describes what to do when encountering common problems:
PROBLEM POSSIBLE CAUSE(S) RECOMMENDATION
SpO
2
NUMBER FLASHES
Saturation alarm limit
exceeded.
Assess/address patient condition
Re-set alarm limits if indicated.
SENSOR OFF MESSAGE
Sensor not connected to
patient properly.
Sensor is damaged.
Properly reapply the sensor on the
patient and reconnect the sensor to
the unit or patient cable.
If the sensor is damaged, replace
the sensor.
NO SENSOR MESSAGE
Sensor is disconnected
from patient cable.
Sensor connected upside
down into patient cable.
Check to see if the sensor LED is
flashing. Disconnect and reconnect
the sensor. If the LED fails to oper-
ate, replace the sensor.
LOW PERFUSION
Improper sensor type.
Poorly perfused site.
Sensor is too tight.
A disorder such as hypo-
thermia, vasoconstriction,
hypovolemia, peripheral
vascular disease or ane-
mia.
Sensor is damaged.
Verify proper sensor and sensor size
for the patient.
Check and see if blood flow to the
site is restricted. Be sure that the
sensor is not on too tight.
Set unit to MAX sensitivity.
Warm the patient or sensor site.
Move sensor to better perfused site.
Vista de pagina 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 55 56

Comentarios a estos manuales

Sin comentarios