3-2 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 3-3
3
setup
Monitor Setup
The Radical-7 Pulse CO-Oximeter maintains three types of default values which will
automatically revert to after a power cycle:
■ Factory – these options are restored to factory set values
■ Custom – these settings can be changed by the user and retained through the
power cycle.
■ Adult / Neo – these settings can be selected to revert to factory or hospital-defi ned
values (for Adult or Neonatal) after a power cycle.
FACTORY SET DEFAULTS
The following outlines the Radical-7 Pulse CO-Oximeter option settings that can be
changed by the user but will revert back to after a power cycle.
OPTION DEFAULT SETTING
LCD SCREEN ILLUMINATION
AC Power
Battery Power
Set to maximum, level 4
Set to minimum, level 1
SENSITIVITY
Set to APOD mode
CUSTOM (USER) DEFINED SETTINGS
This mode is indicated by “Mode Custom” on the Alarms menu.
The following table outlines the options that may be changed by the user and the Radical-7
Pulse CO-Oximeter will remember after a power cycle. The default settings listed below
are those set at the factory and will revert back to if the user chooses to recall the factory
settings. (see Section 4, Display).
OPTION FACTORY DEFAULT SETTING CONFIGURABLE SETTING
DISPLAY VIEW
Set to Pleth + SigIQ
Pleth + SigIQ, Pleth Only,
Numbers
CONTRAST
Set to 32 0 to 63
AVERAGING TIME
Set to 8
2, 4, 8, 10, 12, 14, or 16
seconds
FASTS AT
Set to No Yes/No
HOME USE
Set to No Yes/No
INTERFACE ALARMS
Set to Yes Yes/No
SATSHARE NUMBERS
Set to Yes Yes/No
POWER SAVE
Set to No Yes/No
DATE FORMAT
Set to mm/dd/yy mm/dd/yy and dd/mm/yy
TIME FORMAT
Set to 12 hr. 12 and 24 hour
LANGUAGE
Set to English
See Section 4, Display for all
settings
ANALOG OUTPUT
Set to Analog 1: SpO
2
0-100%
See Section 4, Display for all
settings
Set to Analog 2: Pulse Rate
SERIAL OUTPUT PORT
MODE
Set to Binary
ASCII 1, ASCII 2, Binary,
Philips Vuelink, Spacelabs
Flexport
ALARM VOLUME
Set to Level 3 Level 1 to 4
INITIAL BATTERY CHARGING
Before use, the Radical-7 Pulse CO-Oximeter Handheld battery and the optional
Docking Station battery needs to be fully charged.
To charge the batteries:
1. Attach the Handheld unit to the Docking Station.
2. Plug in the AC power cord to power entry module. Make sure it is securely plugged
in.
3. Plug the AC power cord into an AC power source.
4. Verify that the batteries are charging.
The battery charging LED indicators on the Docking Station flash prior to charging and
remain illuminated while the batteries are charging.
Refer to Section 9, Battery Operation and Maintenance, for proper battery charging.
INITIAL INSTALLATION
Place the Docking Station on a stable hard flat surface near the patient. Always place the
Radical-7 Pulse CO-Oximeter unit on a dry surface. Maintain a minimum of 3 cm (1 inch)
free space around the Radical-7 Pulse CO-Oximeter Standalone unit. Make sure that the
Radical-7 loudspeaker is not covered to avoid a muffled alarm sound.
The Radical-7 Pulse CO-Oximeter Handheld, Docking Station or Standalone should not
be operated outside the following environmental conditions:
OPERATING ENVIRONMENTAL CONDITIONS
TEMPERATURE
+5°C to +40°C, +41°F to +104°F
HUMIDITY
5% to 95%, non-condensing
OPERATING ALTITUDE
1060 mbar to 500 mbar pressure
-1000 ft to 18,000 ft (-304 m to 5,486 m)
Configure the unit for your regional power line frequency (50 or 60 hz) if needed. Default is
60 hz (standard for the United States). See Section 4, Operation, Config.
CAUTION: THE UNIT MUST BE CONFIGURED TO MATCH YOUR LOCAL POWER
LINE FREQUENCY TO ALLOW FOR THE CANCELLATION OF NOISE
INTRODUCED BY FLUORESCENT LIGHTS AND OTHER SOURCES.