4-30 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 5-15-1
5
alarms and messages
g
System Messages
The following chart alphabetically lists all system messages displayed on the LCD screen,
the cause of the message and the action(s) to be taken.
The operator should become thoroughly familiar with this information before using the
oximeter for patient monitoring.
MESSAGE POSSIBLE CAUSE(S) RECOMMENDATION
AMBIENT LIGHT
Too much light on patient
(sensor).
• Inadequate tissue cover-
ing sensor detector.
• Remove or reduce lighting.
• Cover sensor from light.
• Reposition sensor.
DEFECTIVE CABLE
• Oximeter cannot iden-
tify the connected cable
or the cable has failed.
• Inoperative or faulty cable
• Replace cable.
• Refer to the Directions for Use of the cable
being used.
INCOMPATIBLE
SENSOR
• Not a proper Masimo
sensor.
• Replace with a proper Masimo sensor.
• Refer to Section 8.
INVALID SENSOR
• Oximeter cannot
identify the connected
sensor.
• Broken sensor cable wire or inoperative
LEDs or faulty detector.
The sensor has failed.
• Replace sensor.
• Refer to the instructions for the sensor
being used.
LOW BATTERY
• Battery charge is low.
• Charge battery by placing the Radical-7
Handheld into the Docking Station and
powering the unit with AC line power.
• Replace battery if necessary.
LOW PERFUSION
• Signal too small.
• Move sensor to better perfused site.
• Refer to Section 4, Low Perfusion.
LOW SIGNAL IQ
• Low signal quality.
• Ensure proper sensor application.
• Move sensor to a better perfused site.
• Refer to Section 4, Signal IQ.
LOW SpCO CONF
SpCO measurement
reading is obscured.
• Ensure proper sensor application.
• Check sensor to see if it is working
properly.
• If not, replace the sensor.
LOW SpMet CONF
• SpMet measurement
reading is obscured.
• Ensure proper sensor application. Check
sensor to see if it is working properly.
• If not, replace the sensor
SPEAKER FAILURE
• Unit requires service
• Contact Masimo Tech Support.
• Refer to Section 9 Service and repair.
NO CABLE
• Cable not attached or
not fully inserted into
the connector.
• Disconnect and reconnect cable into
connector.
operation
4
To return from SatShare operation to normal standalone operation, simply disconnect
the SatShare cable from the patient monitor or disconnect the SatShare cable from the
SatShare connector on the back of the Radical-7 Pulse CO-Oximeter.
Home Mode Operation
The Radical-7 Pulse CO-Oximeter can be placed into the Home Mode to protect unquali-
fied users from changing the Radical-7 Pulse CO-Oximeter alarm settings and operation.
Entering a password does not automatically reset the Radical-7 Pulse CO-Oximeter to the
Normal operating mode. In the Home Mode, a password is required to access the menu
system and the touch key control buttons and icons.
NOTE: When the Radical-7 Pulse CO-Oximeter is set to operate in the Home mode the
default values that the Radical-7 reverts to after a power cycle are set according to
Section 3, Monitor Setup, with the exception of the Alarm Silence setting, which is
set to the pre-power down setting.
Password Operation
The Radical-7 Pulse CO-Oximeter password is 2-3-1. To enter the password use the touch-
key control buttons to the right or bottom of the LCD display and press the buttons in the
sequence shown in the following figure:
XX X
3
Then press
3 3
First press
2 2 2
1 1
Finally press
1