3-8 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 4-14-1
4
operation
Introduction
To operate the Radical-7 Pulse CO-Oximeter effectively, the device must be set up prop-
erly and the operator must:
Know how the Pulse CO-Oximeter derives its readings (see Section 1, Pulse CO-
Oximetry).
Be familiar with its controls, components and operation.
Understand its status and alarm messages (see Section 5, Alarm and Messages
and Section 6, Troubleshooting).
Basic operation
GENERAL SETUP AND USE
1. Inspect the Pulse CO-Oximeter case for damage.
2.
Connect a patient cable or a direct connect sensor to the Patient Cable Connector of
the Radical-7 Pulse CO-Oximeter. Make sure it is a firm connection and the cable is
not twisted, sliced or frayed.
3. If utilizing the Standalone setup, ensure that the power cord is plugged into the Power
Cable Connector of the Docking Station and into the AC power.
4.
If utilizing a patient cable, select a sensor that is compatible with the oximeter and the
patient before connecting it to the patient cable. See section 8, Sensors and Patient
Cables. If using a single patient adhesive or disposable sensor, check that the emitter
(red light) and the photodetector are properly aligned. Remove any substances that
may interfere with the transmission of light between the sensor’s light source and
photodetector.
5.
Refer to the Directions for Use of the sensor before attaching the sensor to the
patient.
6. With a single patient adhesive or disposable sensor, connect the sensor to the patient
cable with the logos lining up; make sure it is a firm connection.
7. Press the Power/Standby button to turn the Pulse CO-Oximeter on.
8. Make sure the display window is free of alarm and system failure messages (see
Section 5, Alarm Identification).
9. On the display, verify:
The high and low alarm limits for SpO
2
, SpMet*, SpCO* and pulse rate.
The readings for SpO
2
, SpMet*, SpCO*, pulse rate, perfusion index and PVI.
NOTE: “- - - “ will remain lit on the numeric display until the SpO
2
, SpMet*, SpCO*, pulse
rate and PI readings have stabilized (less than 15 seconds for SpO
2
and up to 25 seconds
for SpMet* and SpCO*).
* If these parameters are installed in the unit.
RadNet Setup
CAUTIONS:
AFTER CONNECTING THE SERIAL CABLE TO THE SERIAL PORT, SATSHARE
MONITORING WILL NOT BE SUPPORTED.
SpMet AND SpCO CANNOT BE DISPLAYED ON THE MULTIPARAMETER MONI-
TOR WITH RADNET.
NOTE: Confirm that the Radical Docking Station is RadNet Ready before preceeding.
This is done by pressing the Menu button then selecting Output. Confirm that
the “D-Station” software is version R4.5.0.1 or higher.
1. Select the ASCII 2 selection from the Serial options on the Output Menu on the
Radical-7 Pulse CO-Oximeter.
2. Connect one end of the serial cable to the Serial Output connector on the back of
the Docking Station.
3. Connect the other end of the serial cable to the RadNet Interface Module connec-
tor.
4. Turn the RadNet Interface Module on.
5. With a properly configured RadNet Interface Module, the Radical-7 will automati-
cally display the pleth and SIQ waveforms as well as the SpO
2
and Pulse Rate
parameters on the screen at the RadNet Central Station.
6. The Radical-7 Pulse CO-Oximeter can be set up to audibly indicate all patient
alarms while communicating with the RadNet Interface module. Use the Interface
Alarms setting in the General menu to enable and disable audible alarms on the
Radical-7 Pulse CO-Oximeter.
CAUTION: ENSURE THAT THE RADICAL-7 PULSE CO-OXIMETER HANDHELD
REMAINS IN THE RADNET READY DOCKING STATION WHEN CONNECT-
ED TO THE RADNET CENTRAL MONITORING SYSTEM. REMOVING THE
RADICAL-7 PULSE CO-OXIMETER HANDHELD FROM THE RADNET
READY DOCKING STATION WILL CAUSE LOSS OF COMMUNICATION TO
THE RADNET CENTRAL MONITORING STATION.
CAUTION: WHEN THE RADICAL-7 IS PLACED IN ALL MUTE, THE PATIENT ALARMS
WILL NOT AUDIBLY SOUND ON THE RADICAL-7 OR THE RADNET
CENTRAL STATION. THE RADNET CENTRAL STATION WILL DISPLAY A
VISUAL ALARM.
setup
3