4-16 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 4-17
4
operation
Display continued
MENU ITEMS DESCRIPTION
TREND
(cont.)
Perfusion Index (PI)
Pleth Variability Index (PVI)
Methemoglobin (MET)
Carboxyhemoglobin (CO)
NOTE: These parameters are viewed when utilizing the Trend Display
function and not the Trend Graph function.
CONTRAST
Allows the user to set the contrast of the LCD display. Contrast ranges
from 0 to 63.
NOTE: The contrast can also be set by pressing and holding the
Backlight/ Contrast Button on the front panel.
LANGUAGE
Allows the user to select from the following languages to be displayed
on the screen:
French
Spanish
German
Italian
Portuguese
Danish
Dutch
Swedish
VERTICAL LAYOUT
DEFAULT
Displays the alarm parameters in vertical sequence of SpO
2
, SpMet,
SpCO, BPM, PVI and PI.
TRADITIONAL
Displays the alarm parameters in vertical sequence of SpO2, BPM, PI,
SpMet, SpCO and
PVI.
DEFAULT
Allows the user to select user mode or reset the settings to factory de-
faults.
NOTE: A password is required to access these menu options. Refer to
Section 4 Password Operation
LOW % SpO
2
LIMIT
Allows the qualifi ed user to set a custom default minimum low SpO
2
limit. When set, it will be the lowest value that the low SpO
2
alarm limit
can be set to. For example, if the limit is set to 85%, then it cannot be set
lower than 85% through the main Alarm menu. The unit will return to this
setting after a power cycle.
SAVE LAST
This allows the user to either use Custom setting or Adult/Neonatal set-
tings. Select “Yes” to use custom settings.
SAVE AS ADULT
Store current settings as Adult default setting.
SAVE AS NEO
Store current settings as Neonatal default setting.
RESTORE FACTORY
Recall factory setting for Custom, Adult and Neonatal.
Display
MENU ITEMS DESCRIPTION
VIEW
Five views are available: PVI Pleth and Signal IQ, PVI Pleth, Pleth and
Signal IQ, Numbers and Pleth Only.
PVI PLETH + SIGNAL IQ / PLETH + SIGNAL IQ
Shows the SpO
2
and pulse rate numbers on the left or top of the screen.
The PVI plethysmograph-Signal IQ and plethysmograph-Signal IQ wave-
forms are on the right, two-thirds or bottom of the screen. The screen
also indicates the signal strength of the measured signal as a perfusion
index (PI). The PI is calculated as the relation of arterial pulsatile signal
to the non-pulsatile signal component. The percentage measurements of
methemoglobin (SpMet) and carboxyhemoglobin (SpCO) are displayed
in the middle upper third of the screen above the PI measurement. The
PVI measurement is displayed under the SpCO measurement.
PVI PLETH / PLETH ONLY
Shows the SpO
2
and pulse rate numbers on the left or top of the screen.
The PVI plethysmograph and plethysmograph waveform are on the
right, two-thirds or bottom of the screen. The screen also indicates the
signal strength of the measured signal as a perfusion index (PI). The PI
is calculated as the relation of arterial pulsatile signal to the non-pulsa-
tile signal component.The percentage measurements of methemoglobin
(SpMet) and carboxyhemoglobin (SpCO) are displayed in the middle
upper third of the screen above the PI measurement. The PVI measure-
ment is displayed under the SpCO measurement.
NUMBERS
Shows the SpO
2
and pulse rate numbers and the signal IQ in the form
of a pulse bar on the screen. The Screen also indicates the signal
strength of the measured signal as a perfusion index (PI). The PI is
calculated as the relation of arterial pulsatile signal to the non-pulsatile
signal component. The percentage measurements of methemoglobin
(SpMet) and carboxyhemoglobin (SpCO) are displayed in the middle
of the screen in line with the PI measurement. The PVI measurement is
displayed under the PI measurement.
TREND
This allows the user to select and view the Trend data between:
SpO
2
and pulse rate (%SpO
2
+ BPM)
SpO
2
and perfusion index (%SpO
2
+ PI)
SpO
2
and methemoglobin (%SpO
2
+ MET)
SpO
2
and carboxyhemoglobin (%SpO
2
+ CO)
SpO
2
and Pleth Variability Index (%SpO
2
+ PVI)
Pulse rate and perfusion index (BPM + PI)
Pulse rate and methemoglobin (BPM + MET)
Pulse rate and carboxyhemoglobin (BPM + CO)
Pulse rate and Pleth Variability Index (BPM + PVI)
Perfusion index and methemoglobin (PI + MET)
Perfusion index and carboxyhemoglobin (PI + CO)
Perfusion index and Pleth Variability Index (PI + PVI)
Methemoglobin and carboxyhemoglobin (MET + CO)
Methemoglobin and Pleth Variability Index (MET + PVI)
Carboxyhemoglobin and Pleth Variability Index (CO + PVI)
%SpO
2
Pulse Rate (BPM)
operation
4-16
4