4-14 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 4-154-15
4
operation
Rotate Screen
MENU ITEMS DESCRIPTION
LANDSCAPE 1
The viewing screen will be in a horizontal fashion when the Radical-7 is
in the upright, horizontal position.
VERTICAL 1
The viewing screen will be in a vertical fashion when the Radical-7 is in
the upright, vertical position.
LANDSCAPE 2
The viewing screen will be in an inverted horizontal fashion (rotated
180
o
) when the Radical-7 is in the upright, horizontal position.
VERTICAL 2
The viewing screen will be in an inverted vertical fashion (rotated 180o)
when the Radical-7 is in the upright, vertical position.
Alarms continued
MENU ITEMS DESCRIPTION
VOLUME
This menu allows the user to set the alarm volume. Four levels are avail-
able: level 1 being the softest and level 4 being the loudest. The device
retains the Alarm Volume setting upon a power cycle.
NOTE: For home use, set the alarm level to level 4.
DELAY
This menu allows the users to set an audible saturation delay. The delay
can be set to either 0, 5, 10 or 15 seconds. The delay setting only affects
saturation alarm indications.
NOTE: In “Custom” mode (see Section 4-Operation) the unit will retain
the Alarm Delay setting after a power cycle. In “Neo” or “Adult” mode
(see Section 4-Operation) the unit will reset the Alarm Delay to the hos-
pital specifi ed setting after a power cycle.
MODE
The Radical-7 stores three types of modes: Adult, Neonatal or Custom
limits. Adult and Neonatal must be initially set and enabled (via pass-
word protected screen) before they can be selected.
CUSTOM
Any changes to settings on the Alarm menu will be retained after a
power cycle.*
ADULT
Any changes to settings on the Alarm menu will be reset to pre-defi ned
Adult defaults after a power cycle.†
NEO
Any changes to settings on the Alarm menu will be reset to pre-defi ned
Neonatal defaults after a power cycle.†
TYPES
SpO
2
(HIGH)
SpO
2
(LOW)
PULSE RATE
(HIGH)
PULSE RATE
(LOW)
ADULT
LIMITS†
Off 90% 140 BPM 50 BPM
NEONATAL
LIMITS†
100% 90% 180 BPM 100 BPM
CUSTOM
LIMITS*
Off* 90%* 140 BPM* 50 BPM*
NOTE: Limits are set at the factory to the values listed in this table.
*Once Custom values are changed by the user, they will be retained
after a power cycle.
†Adult and Neo settings can be changed (via a password protected
screen, see Section 4 under Password Operation) to specifi c hospital/
unit requirements. If the settings are changed, then any values changed
by the user will be returned to the unit's default values after it is powered
down.
operation
4-14
4