4-6 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 4-74-7
4
operation
Touch Key Control Button and Icons
The touch key control buttons are the four dark grey control buttons to the right of the
Handheld display. To select a touch key icon, press and release the dark gray control button
to the right of the icon.
On the Radical-7 Pulse CO-Oximeter display, four icons are shown on the right side or
bottom of the LCD display.
Traditional User Interface
FIRST PAGE
NEXT MENU
PAG E
Press the Next Menu Page button to access the second page of
selections.
Press and hold the button for 8 seconds to toggle between
Traditional User Interface and the Simplified User Interface. Enter
password when prompted. The Radical-7 will retain this setting
after a power cycle.
MENU ACCESS
Press the Menu Access button to enter the main menu.
APOD
MAX
SENSITIVITY
Press the Sensitivity button to toggle between the Normal, APOD
and Maximum Sensitivity modes. Use the Normal Sensitivity set-
ting for typical monitoring purposes. Use the APOD setting where
there is a high probability of the sensor becoming detached. Use
the Maximum Sensitivity setting for patients with low perfusion or
when the low perfusion message is displayed on the screen in
APOD or normal sensitivity mode. The default is APOD.
CAUTION: WHEN USING THE MAXIMUM SENSITIVITY SET-
TING, THE PERFORMANCE OF THE SENSOR OFF DETEC-
TION MAY BE COMPROMISED.
NOTE: In “Custom” mode the unit will remain in Normal or APOD
setting after a power cycle. Maximum Sensitivity will automatically
reset to Normal Sensitivity after a power cycle.
In “Neo” or “Adult” mode the unit will reset the sensitivity to the
hospital specified setting (Normal or APOD) after a power cycle.
T
REND
GRAPH
TREND GRAPH
Press the Trend Toggle button to alternate between SpO
2
, BPM,
SpMet, SpCO, PI and PVI Quick Trend displays.
SECOND PAGE
NEXT MENU
PAG E
Press the Next Menu Page button to access the first page of
selections.
Press and hold the button for 8 seconds to toggle between
Traditional User Interface and the Simplified User Interface. Enter
password when prompted. The Radical-7 will retain this setting
after a power cycle.
TREND DISPLAY
Press the Trend Display button to show the trend data on the dis-
play.
INCREASE
LOUDNESS
Press the Increase Loudness button to increase the volume of the
pulse beep.
Seven levels of volume exist
DECREASE
LOUDNESS
Press the Decrease Loudness button to decrease the volume of
the pulse beep.
If artificial nails or excessive fingernail polish are present, select another site or
remove the polish/artificial nails.
If possible, ensure that the sensor is placed in a location with low ambient light.
Although the Radical-7 with integrated Masimo Rainbow SET technology has sig-
nificant immunity to ambient light, excessive ambient light may cause readings to be
incorrect.
CAUTION: IF ANY MEASUREMENT SEEMS QUESTIONABLE, FIRST CHECK THE
PATIENT’S VITAL SIGNS BY ALTERNATE MEANS AND THEN CHECK THE
PULSE CO-OXIMETER FOR PROPER FUNCTIONING.
SENSITIVITY
The Radical-7 Pulse CO-Oximeter is equipped with 3 different sensitivity modes. Each
mode allows the clinician to change the sensitivity settings of the device to meet the
increased demands of the patient’s physiological condition or enable it to work during
periods of low perfusion and/or motion. They are as follows:
Normal SensitivityThis is the recommended mode for patients that are experienc-
ing some compromise in blood flow or perfusion. It is advisable for care areas where
patients are observed frequently, such as ICU’s.
Adaptive Probe Off Detection (APOD)This is the recommended start-up monitor-
ing mode for most patients with acceptable perfusion or where a more robust sensor
off detection is desired. It is the suggested mode for care areas where patients
are not visually monitored continuously. This mode delivers enhanced protection
against erroneous pulse rate and arterial oxygen saturation readings when a sensor
becomes inadvertently detached from a patient.
Maximum Sensitivity (MAX) - This mode is recommended for patients with low
perfusion or when the low perfusion or low signal quality message is displayed on
the screen in APOD or normal sensitivity mode. This mode is not recommended for
care areas where patients are not monitored visually, such as general wards. It is
designed to interpret and display data at the measuring site when the signal may be
weak due to decreased perfusion. When a sensor becomes detached from a patient,
it will have compromised protection against erroneous pulse rate and arterial satura-
tion readings.
CAUTION: WHEN USING THE MAXIMUM SENSITIVITY SETTING, THE PERFOR-
MANCE OF THE SENSOR OFF DETECTION MAY BE COMPROMISED.
IF THE UNIT IS IN THIS SETTING AND THE SENSOR BECOMES DIS-
LODGED FROM THE PATIENT, THE POTENTIAL FOR FALSE READINGS
MAY OCCUR DUE TO ENVIRONMENTAL ‘NOISE’ SUCH AS LIGHT, VIBRA-
TION AND EXCESSIVE AIR MOVEMENT.