alarms and messages
3D Alarm System Option
This section applies only if you have purchased the optional 3D Alarm System. Contact
Masimo if you are interested in purchasing this option.
Purpose
The purpose of this section is to familiarize clinicians with the optional 3D Alarm System of
the Radical-7 Pulse CO-Oximeter.
3D Alarm System Overview
The Radical-7 Pulse CO-Oximeter includes user-selectable High and Low alarm limits for
SpO
2
and pulse rate to provide audible and visual indication of specific levels of these vital
signs that the clinician has determined merits their attention as described in Section 4 of
this manual. The 3D Alarm System enables clinicians to be alerted to changes in multiple
interacting factors to provide an additional level of vigilance and flexibility to manage their
patients.
The following is a summary of each of the 3D Alarm System features:
DESAT INDEX ALARM
The Desat Index Alarm is a user-selectable feature which allows a clinician to request an
audible and visual alarm if a patient experiences a specified number of desaturations over
a specific period of time.
NOTE: The alarms in this option are considered to be Medium priority.
PERFUSION INDEX (PI) DELTA ALARM
The PI Delta Alarm is a user-selectable feature which allows a clinician to request an
audible and visual alarm if perfusion at the monitored site decreases by a specified level
(delta) over a specific period of time.
NOTE: The alarms in this option are considered to be Medium priority.
5-2
alarms and messages
5
MESSAGE POSSIBLE CAUSE(S) RECOMMENDATION
NO SENSOR
• Sensor not fully
inserted into the con-
nector.
• Maybe an incorrect sensor, or a defective
sensor or cable.
• Insert sensor into connector.
• Disconnect and reconnect sensor.
• Refer to the instructions for the sensor
being used.
• Unit is searching for
patient’s pulse.
• Disconnect and reconnect the sensor into
the Patient Cable Connector.
PULSE SEARCH
• Unit is searching for
patient’s pulse.
• If values are not displayed within 30 sec-
onds, disconnect and reconnect sensor.
• If pulse search continues, remove sensor
and replace on a better perfused site.
SENSOR
CALIBRATING
• Unit is checking the
sensor for proper
functioning and perfor-
mance.
• If values are not displayed within 30 sec-
onds, disconnect and reconnect sensor.
• If values are still not displayed, replace
with a new sensor.
SENSOR OFF
• Sensor off patient.
• Disconnect and reconnect sensor.
• Reattach sensor.
SERVICE RE-
QUIRED*
• Internal Failure. • Unit requires service.
UNRECOGNIZED
CABLE
• Not a proper cable.
• Replace with a proper cable. Refer to
Section 8.
*The SERVICE REQUIRED message fills the entire display. This is a numeric error code.
Contact Masimo for service.