Masimo Radical Signal Extraction Manual del operador Pagina 34

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 33
4-26 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 4-27
4
operation
MENU ITEMS DESCRIPTION
TREND ACTION
SERIAL DUMP
To send all the data that is stored in trend memory to the serial port, se-
lect the Serial Dump option. Use this option to communicate the stored
data set to trend graphing software applications.
ANALOG DUMP
To send all the data that is stored in the trend memory to the analog
output select the Analog Dump option. Use this option to print the trend
information on an analog chart recorder.
PRINT
To print the trend data that is shown in the Trend View select the Print
option. The trend data is fi rst printed in histogram format, followed by a
table of data that shows the time and date stamp of a trend record, and
the SpO
2
, pulse rate, SpMet, SpCO, PI and PVI measurement. Each
trend record is printed on a single line followed by a carriage return and
line feed character.
TREND PERIOD
The Trend period setting determines how often a set of SpO
2
, pulse rate,
SpMet, SpCO and PI data points is stored in trend memory. A setting of
2, for example, sets the Radical-7 Pulse CO-Oximeter to store one set
of SpO
2
, pulse rate, SpMet, SpCO, PI and PVI measurements every 2
seconds, resulting in a minimum trend capacity of 72 hours. A setting
of 10, for example sets the Radical-7 Pulse CO-Oximeter to store one
set of data points every 10 seconds, resulting in a typical trend storage
capacity of 18 days.
NOTE: Since the Radical-7 Pulse CO-Oximeter employs a sophisticated data compression
scheme, the actual trend capacity is dependent on the type of data that is collected.
HISTOGRAM
The figure below is an example of the display output of the Histogram function. The
Histogram displays the trend data of selected parameters as numerical percentages shown
in specific ranges.
%SpO
2
PI
PI
%SpO
2
000
0.0
0.0 0.0
1 Hr
4/11/06
Histogram
14:26:38 15:26:38
96-100 0
0
0
0
0
0
0
0
0
0
91-95
86-90
16.1-20.0
%
%
81-85
1-80
12.1-16.0
8.1-12.0
4.1-8.0
0.1-4.0
Masimo SET
98
100
140
50
85
%SpO
2
BPM
71
Signal Extraction Pulse CO-Oximeter
operation
4
TREND SETUP
This menu allows the user to set the default trend settings or download the trend data. The
default settings are used to scale the trend graphs when the trend data button, located on
the main display, is accessed.
MENU ITEMS DESCRIPTION
%SpO
2
MAX SCALE
Sets the maximum scale of the SpO
2
trend graph.
%SpO
2
MIN SCALE
Sets the minimum scale of the SpO
2
trend graph.
PR MAX SCALE (BPM)
Sets the maximum scale of the pulse rate trend graph.
PR MIN SCALE (BPM)
Sets the minimum scale of the pulse rate trend graph.
PI MAX SCALE
Sets the maximum scale of the PI trend graph.
PI MIN SCALE
Sets the minimum scale of the PI trend graph.
%SpMet MAX SCALE
Sets the maximum scale of the SpMet trend graph.
%SpMet MIN SCALE
Sets the minimum scale of the SpMet trend graph.
%SpCO MAX SCALE
Sets the maximum scale of the SpCO trend graph.
%SpCO MIN SCALE
Sets the minimum scale of the SpCO trend graph.
PVI MAX SCALE
Sets the maximum scale of the PVI trend graph.
PVI MIN SCALE
Sets the minimum scale of the PVI trend graph.
DEFAULT VIEW
Selects the default time scale of the trend view. This setting only se-
lects the time scale of the trend view when the trend data is initially dis-
played, (i.e. when the trend data is initially accessed). The selections
are 24 hrs, 12 hrs, 8 hrs, 4 hrs, 2 hrs, 1 hr, 30 minutes, 10 minutes, 1
minute and 20 seconds.
Max
Min
Default View
Trend Action
Trend Period
Trend Setup
%SpO
2
100
BPM PI
%SpMet
%SpCO
200
20
100
50 25 0 50
40
0
1 Hr
None
2
PVI
20
0
Masimo SET
98
100
140
50
85
%SpO
2
BPM
71
Signal Extraction Pulse CO-Oximeter
Vista de pagina 33