8-2 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual
8
sensors & patient cables
Masimo SpO
2
Sensors
Before use of any sensor or cable, carefully read the sensor or cable Directions for Use.
Use only Masimo oximetry sensors and cables for SpO
2
measurements. Other oxygen
transducers or sensors may cause improper Rad-8 Pulse Oximeter performance.
Tissue damage can be caused by incorrect application or use of a Masimo sensor, for
example by wrapping the sensor too tightly. Inspect the sensor site as directed in the
sensor Directions for Use to ensure skin integrity and correct positioning and adhesion
of the sensor.
CAUTIONS:
DO NOT USE DAMAGED SENSORS. DO NOT USE A SENSOR WITH EXPOSED
OPTICAL OR ELECTRICAL COMPONENTS. DO NOT IMMERSE THE SENSOR
IN WATER, SOLVENTS, OR CLEANING SOLUTIONS (THE SENSORS AND
CONNECTORS ARE NOT WATERPROOF). DO NOT STERILIZE BY IRRADIATION,
STEAM, OR ETHYLENE OXIDE. SEE THE CLEANING INSTRUCTIONS IN THE
DIRECTIONS FOR USE FOR REUSABLE MASIMO SENSORS.
DO NOT USE DAMAGED PATIENT CABLES. DO NOT IMMERSE THE PATIENT
CABLES IN WATER, SOLVENTS, OR CLEANING SOLUTIONS (THE PATIENT
CABLE CONNECTORS ARE NOT WATERPROOF). DO NOT STERILIZE BY
IRRADIATION, STEAM, OR ETHYLENE OXIDE.
ALL SENSORS AND CABLES ARE DESIGNED FOR USE WITH SPECIFIC
MONITORS. VERIFY THE COMPATIBILIY OF THE MONITOR, CABLE AND
SENSOR BEFORE USE, OTHERWISE PATIENT INJURY CAN RESULT.
DO NOT USE ADDITIONAL TAPE TO WRAP SENSOR.
RED DIRECT CONNECT SENSORS
Masimo Red Direct sensors can be used with the Rad-8 to enable measurement of SpO
2
and pulse rate only. Red Direct Connect sensors will only function with oximeter devices
equipped with Masimo SET technology. Red Direct Connect sensors connect to the device
directly.
SENSOR
Weight
Range
Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
No Motion Motion No Motion Motion Saturation Pulse Rate
DC-3
DC-12
> 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
DCP-3
DCP-12
10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
SENSOR
Weight
Range
Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
No Motion Motion No Motion Motion Saturation Pulse Rate
Red DCI-dc3 > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
Red DCI-dc12 > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
Red DCIP-dc3
10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
Red DCIP-dc12
10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm