4-14 Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual
4
SETUP MENU LEVEL 3 (ENTER BUTTON + DOWN BUTTON MENU SETTINGS)
Setup Menu Level 3 contains advanced parameter/measurement settings.
To access Level 3 parameters/measurements,
hold down the Enter Button and press the
Down Button for 5 seconds. After entering menu Level 3, use the Up or Down button to
move between settings. Press the Enter Button to save new settings and move to the next
menu. The user cycles through the menu options by continuing to press the Enter Button.
AVERAGING TIME
BUTTONS SETTINGS INSTRUCTIONS
Hold down Enter
Button and Down
Button for 5 seconds.
+
Averaging Time
8 seconds
(default, except in
Sleep Mode)
Use the Up or Down Button
to move between settings
AND
press the Enter Button to
accept the setting and move
to the next menu option
OR
let the device time out for
10 seconds to exit without
saving the new setting.
NOTE: If the device is in the
Sleep Mode, averaging time
automatically reverts to 2
seconds.
Press Up Once
Averaging Time
10 seconds
Press Up 2x
Averaging Time
12 seconds
Press Up 3x
Averaging Time
14 seconds
Press Up 4x
Averaging Time
16 seconds
Press Up 5x
Averaging Time
2 seconds
Press Up 6x
Averaging Time
4 seconds
operation