Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-3
4
operation
FACTORY DEFAULT AND USER CONFIGURABLE SETTINGS
OPTION
FACTORY DEFAULT
SETTINGS
USER CONFIGURABLE
DEFAULT SETTINGS
USER CONFIGURABLE
DEFAULT SETTINGS
REFERENCE
SpO
2
high alarm
limit*
"---" Off 2 to 99%
Section 4
Setup Menu Level 1
SpO
2
low alarm
limit*
90% 1 to 99%
Pulse rate high
alarm limit*
140 BPM 35 to 235 BPM
Pulse rate low
alarm limit*
50 BPM
30 to 230 BPM
LED Brightness Level 3 Levels 1 thru 4
Alarm Volume Level 3 (70 dB min) Levels 1 thru 4
Section 4
Setup Menu Level 2
Alarm Silence 120 seconds 30, 60, 90, or 120 seconds
Alarm Delay 5 seconds 0, 5, 10, or 15 seconds
Clear Trend Clear Trend (No) Clear Trend (Yes/No)
Button Volume Level 2
Levels 1 thru 4
Sensitivity APOD
Max/Normal/APOD
NOTE: MAX sensitivity will
default to APOD after a
power cycle.
FastSat FastSat Off
FastSat (On/Off)
Averaging Time
8 seconds (2 sec. in
Sleep mode)
2, 4, 8, 10, 12, 14 and 16
Section 4
Setup Menu Level 3
Rapid Desat Alarm 5% 0, 5 and 10
Alarm On/Off Alarms active (On)
On/Off or muted with
reminder
Default Settings No Change User Default/Factory Default
SmartTone On/Off
Off (Normal Tone) On (SmartTone)
Year
Internal Clock
Current year
Month Curent Month
Day Current Day
Hour Current Hour
Minute Current Minute
Serial Out ASCII 2 Philips/ASCII 1/ASCII 2
Software Version System Software SET Software
Interface Alarms Alarms On Alarms (On/Off)
Nurse Call Alarms Alarms/SIQ/Alarms and SIQ
Polarity Normal Normal/Invert
Set Mode Standard (Normal)
Standard/Sleep/Home**
Section 4
Setup Menu Level 4
* User configurable settings will change back to factory defaults when power is turned off and on,