Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-23
4
operation
Trend Setup and Use
INTRODUCTION
The Rad-8 can store up to 72 hours of trend data captured at 2 second intervals. The trend
data can then be transferred to a PC for evaluation.
A serial cable is required to connect the Rad-8 to a PC. Patient monitoring is not possible
while trend memory is being transferred to a PC.
Trend data is stored in non-volatile memory, so it is not erased when the device is shut off. A
trend data download is initiated using the TrendCom utility which downloads the trend data
and saves it to an ASCII text (.out) file with an output delimiter option.
NOTE: Rad-8 Serial Output must be set to
ASCII 2 for successful download of trending
data. Refer to the Serial Output settings, Setup Menu Level 3, in this section.
TRENDCOM UTILITY INSTALLATION
Copy the TrendCom utility from the TrendCom CD onto a PC running MS-Windows.
TRENDCOM UTILITY OPERATION
NOTE: During download of trend information, all normal Rad-8 functions are unavailable
and the keypad is locked, except for the power button.
NOTE: If your PC is unable to recognize the USB to Serial port adapter, contact USB to
serial port adapter manufacturer for assistance or support.
1. Turn Rad-8 off if not already off.
2. Connect serial cable to Rad-8 and other end to a com port on the PC.
3. Turn the Rad-8 on.
4. Start the TrendCom utility on the PC.
5. Select Rad-8 from the first pull-down menu.
6. Select the appropriate com port number from the second pull-down menu, if necessary.
7. Select the Output Delimiter Option (Tab, Comma or Space).
8. Select the RETRIEVE TREND button on the TrendCom utility. Select the desired
location and assign a file name for the trend file. Select SAVE.
9. The Rad-8 will display “dat out” while trend data is being transferred. A progress bar
will advance to indicate the status of the download. Larger trend files will take longer
to download. Transfer time is approximately 20 seconds per hour of trend data.
10 When trend data transfer is complete, close TrendCom and disconnect the Rad-8
from the serial cable.
11. Turn the Rad-8 off to exit the trend download mode.
ERASING TREND MEMORY
To erase (clear) the trend memory, set the Clear Trend option to Yes and press the Enter
Button to accept the setting and clear the data from the memory. Refer to the Clear Trend
menu located before this Trend setup and use section.
The Rad-8 continuously trends data. When performing a new study and gathering data
on a new patient, it is highly recommended the "clear function" be utilized in order for the
results to be separate. Turning the Rad-8 off will not erase the trend data.