Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-11
4
operation
SETUP MENU LEVEL 2 (CONTINUED)
ALARM SILENCE
BUTTONS SETTINGS INSTRUCTIONS
Press the Enter Button
again to move to the
next menu.
Alarm Silence
120 seconds
(default)
Use the Up or Down Button
to move between settings and
the Enter Button to accept the
setting and move to the next
menu screen
OR
let the device time out for 10
seconds to exit without saving
the new setting.
Press Up Once
Alarm Silence
30 seconds
Press Up 2x
Alarm Silence
60 seconds
Press Up 3x
Alarm Silence
90 seconds
ALARM DELAY
Alarm delay allows the user to adjust the time in which the audible status indicator will
occur after an alarm condition has been initiated.
BUTTONS SETTINGS INSTRUCTIONS
Press the Enter Button
again to move to the
next menu.
Alarm Delay
5 seconds
(default)
Use the Up Button to move
between settings and the
Enter Button to accept the
setting and move to the next
menu screen
OR
let the device time out for 10
seconds to exit without saving
the new setting.
Press Up Once
Alarm Delay
10 seconds
Press Up 2x
Alarm Delay
15 seconds
Press Up 3x
Alarm Delay
0 seconds