
Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-13
4
operation
SENSITIVITY
BUTTONS SETTINGS INSTRUCTIONS
Press the Enter Button
again to move to the
next menu.
Sensitivity
APOD (APO)
(default)
Use the Up or Down Button
to move between settings
and the Enter Button to
accept the setting and
move to the next menu
screen
OR
let the device time out for
10 seconds to exit without
saving the new setting.
Press Up Once
Sensitivity
NORM (NOr)
Press Up 2x
Sensitivity
MAX
FASTSAT
BUTTONS SETTINGS INSTRUCTIONS
Press the Enter Button
again to move to the
next menu.
FastSat Off
(Default)
Use the Up or Down
Button to move between
settings and the Enter
Button to accept the
setting and move to the
next menu screen
OR
let the device time out for
10 seconds to exit without
saving the new setting.
Press Up Once
FastSat On
SETUP MENU LEVEL 2 (CONTINUED)
Comentarios a estos manuales