Rad-8 Signal Extraction Pulse Oximeter Operator’s Manual 4-25
4
operation
HOME MODE OPERATION
The device can be placed into the Home Mode to protect unqualified users from changing
the alarm settings and operation. Only the following menu and front panel functions are
available: display brightness, pulse beep volume adjustment, and alarm suspend. Alarm
volume is at highest setting. All default and user defined default settings are locked to
their current values when home mode is selected and return to those values after a power
cycle. Upon power up, the Hmm mode will be displayed along with a 10 second display of
parameters. To turn the unit off the power key must be depressed and held for 3 seconds.
The Mode Enter and Up Button held simultaneously for 3 second will put it back into the
special menu to select a different mode.
SLEEP MODE OPERATION
The device can be placed into the Sleep Mode (SLP) to allow the unit to capture normal
and abnormal patient data without triggering the alarms. This mode will blank out the unit
display with the exception of the Battery Level Indicator and the Alarm Silenced Indicator
and disable the alarms even after a power cycle. However, any single key press will bring
the display back for 10 seconds. Upon power up, the SLP mode will be displayed along
with a 10 second display of parameter settings. The Sleep mode automatically defaults to
2 second averaging time. Coming out of the Sleep mode returns the user to orginally set
averaging time. The Mode Enter and Up Button held simultaneously for 3 seconds (select
next (STD), Mode Enter) will put it back into the special menu to exit.
CAUTION: ALARMS ARE DISABLED IN THIS MODE.
NOTE: If the unit is connected to a Patient Safety Net system, there will be no communication
with the Patient Safety Net in Home and Sleep mode.