Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 6-3
6
PROBLEM POSSIBLE CAUSE(S) RECOMMENDATION
PRINT FUNCTION DOES
NOT WORK
Wrong serial cable is used
Make sure a null modem cable is
used.
LED LIGHTS ON LEFT
SIDE OF DOCKING
STATION CONTINUOUSLY
FLASH
Incompatible version of
software on Radical-7
handheld and docking
station.
Upgrade to current software versions.
Match handheld to docking station
with compatible software versions
BATTERY RUN-TIME IS
SIGNIFICANTLY REDUCED
Battery Memory effects.
Use Battery Discharge function as
described in Section 4, Service.
Troubleshooting