9-4 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
9
Service / Maintenance
REPLACING THE BATTERIES
Before installing or removing the battery, make sure the AC power cord is removed and power to
the Pulse CO-Oximeter is turned off.
To replace the Handheld battery, follow these instructions:
1.
Turn the Radical-7 Pulse CO-Oximeter Handheld off and remove the patient cable connection.
Detach the Radical-7 Pulse CO-Oximeter Handheld from the Docking Station (if docked).
2. Loosen the closure screw on the battery compartment door and lift out the battery.
3. Take a new battery, and place it in the compartment.
4. Tighten the closure screw.
5. Place Handheld into Docking Station, turn on line power and charge battery according to
this Section, Battery Operation and Maintenance.
CAUTION: FOLLOW LOCAL GOVERNING GUIDELINES FOR PROPER DISPOSAL OF
INTERNAL BATTERIES. DO NOT INCINERATE
.
WARNING:
THE DOCKING STATION BATTERY SHOULD BE INSTALLED AND/OR
REMOVED FROM DOCKING STATION BY QUALIFIED PERSONNEL ONLY.
REPLACING THE FUSES
Should a power problem blow one or both of the fuses in the power entry module on the rear
panel, the fuse(s) will need to be replaced.
To replace the fuse(s), you will need a flat-blade screwdriver (5mm; 3/16”).
To replace the fuses:
1. Disconnect instrument from AC power.
2. Remove AC power cord from the power entry module at the rear of the docking station.
3. Use the small flat-blade screwdriver and gently pry loose the fuse cover in the left portion
of the power entry module, exposing the fuse holder.
4. Using the small flat-blade screwdriver, gently remove the fuse holder.
5. Note how the fuse(s) are placed in the fuse holder for installation of the new fuse(s).
6.
To remove the fuses from the fuse holder, use the edge of the screwdriver blade to pry against
the bottom of the metal portion of the fuse where it is secured to the glass portion of the fuse.
7. Place the fuse(s) (1 Amp, Metric, fast acting, 5x20mm, 250V) in the fuse holder, properly
orienting the fuse(s).
8. Slide the fuse holder back into the power entry module and press firmly to make sure it is
completely seated.
9. Close the fuse cover and press gently until it seats completely, flush with the back of
the docking station.
10. The instrument is ready to be reconnected to AC power.
NOTE: If the fuses blow shortly after replacement, the instrument requires service.
WARNING:
FIRE HAZARD: TO PROTECT AGAINST FIRE HAZARD, REPLACE ONLY WITH
FUSES OF THE SAME TYPE, CURRENT RATING, AND VOLTAGE RATING.