3-8 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
3
5. In order for the plethysmographic waveform to be displayed on the Spacelabs screen
and for the Spacelabs monitor to indicate the alarm conditions measured by the Pulse
CO-Oximeter, the user must configure the Spacelabs monitor. Refer to the Spacelabs
monitor Operator’s Manual for complete instructions.
6. The Radical-7 Pulse CO-Oximeter can be set up to audibly indicate all patient alarms
while communicating with the Spacelabs Flexport module. Use the Interface Alarms
setting in the General menu to enable and disable audible alarms on the Radical-7
Pulse CO-Oximeter.
Patient SafetyNet/RadNet Setup
CAUTIONS:
SIMULTANEOUS USE OF SATSHARE AND SERIAL PORT IS NOT SUPPORTED.
ONLY SpO
2
AND PULSE RATE CAN BE DISPLAYED ON THE MULTIPARAMETER
MONITOR WITH PATIENT SAFETYNET AND RADNET.
NOTE: Confirm that the Radical Docking Station is Patient SafetyNet and RadNet Ready
before proceeding. This is done by pressing the Menu button then selecting About.
Confirm that the “D-Station” software displays (RadNet) next to the version number.
1. Select the ASCII 2 selection from the Serial options on the Output Menu on the
Radical-7 Pulse CO-Oximeter.
2. Connect one end of the serial cable to the Serial Output connector on the back of the
Docking Station.
3. Connect the other end of the serial cable to the Patient SafetyNet wireless radio or
RadNet Interface Module connector.
4. Turn the Patient SafetyNet wireless radio or RadNet Interface Module on.
5. With a properly configured, the Patient SafetyNet wireless radio or RadNet Interface
Module, the Radical-7 will automatically display the plethysmographic and SIQ
waveforms as well as the installed numeric parameters on the Patient SafetyNet Detail
View and on the screen at the RadNet Central Station.
CAUTION: ENSURE THAT THE RADICAL-7 PULSE CO-OXIMETER HANDHELD
REMAINS IN THE RADNET READY DOCKING STATION WHEN CONNECTED
TO THE MONITORING SYSTEM. REMOVING THE RADICAL-7 PULSE
CO-OXIMETER HANDHELD FROM THE RADNET READY DOCKING
STATION WILL CAUSE LOSS OF COMMUNICATION TO THE PATIENT
SAFETYNET OR
RADNET CENTRAL MONITORING STATION.
CAUTION: WHEN THE RADICAL-7 IS PLACED IN ALL MUTE, THE PATIENT ALARMS
WILL NOT AUDIBLY SOUND ON THE RADICAL-7, THE PATIENT SAFETYNET
OR THE RADNET CENTRAL STATION. THE RADNET CENTRAL STATION
WILL DISPLAY A VISUAL ALARM.
Setup