9-4 Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual 9-5
REPLACING THE BATTERIES
■ Before installing or removing the battery, make sure the AC power cord is removed
and power to the Pulse CO-Oximeter is turned off.
To replace the Handheld battery, follow these instructions:
1. Turn the Radical-7 Pulse CO-Oximeter Handheld off and remove the patient cable
connection. Detach the Radical-7 Pulse CO-Oximeter Handheld from the Docking
Station (if docked).
2. Loosen the closure screw on the battery compartment door and lift out the battery.
3. Take a new battery, and place it in the compartment.
4. Tighten the closure screw.
5. Place Handheld into Docking Station, turn on line power and charge battery according
to this Section, Battery Operation and Maintenance.
CAUTION: FOLLOW LOCAL GOVERNING GUIDELINES FOR PROPER DISPOSAL
OF INTERNAL BATTERIES. DO NOT INCINERATE.
WARNING: THE DOCKING STATION BATTERY SHOULD BE INSTALLED AND/
OR REMOVED FROM DOCKING STATION BY QUALIFIED
PERSONNEL ONLY.
REPLACING THE FUSES
Should a power problem blow one or both of the fuses in the power entry module on the
rear panel, the fuse(s) will need to be replaced.
To replace the fuse(s), you will need a flat-blade screwdriver (5mm; 3/16”).
To replace the fuses:
1. Disconnect unit from AC power.
2. Remove AC power cord from the power entry module at the rear of the docking sta-
tion.
3. Use the small flat-blade screwdriver and gently pry loose the fuse cover in the left
portion of the power entry module, exposing the fuse holder.
4. Using the small flat-blade screwdriver, gently pry out the fuse holder.
5. Note how the fuse(s) are placed in the fuse holder for installation of the new fuse(s).
6. To remove the fuses from the fuse holder, use the edge of the screwdriver blade to
pry against the bottom of the metal portion of the fuse where it is secured to the glass
portion of the fuse.
7. Place the fuse(s) (1 Amp, Metric, fast acting, 5x20mm, 250V) in the fuse holder,
properly orienting the fuse(s).
8. Slide the fuse holder back into the power entry module and press firmly to make sure
it is completely seated.
9. Close the fuse cover and press gently until it seats completely, flush with the back of
the docking station.
10. The unit is ready to be reconnected to AC power.
NOTE: If the fuses blow shortly after replacement, the unit requires service.
WARNING: FIRE HAZARD: TO PROTECT AGAINST FIRE HAZARD, REPLACE
ONLY WITH FUSES OF THE SAME TYPE, CURRENT RATING, AND
VOLTAGE RATING.
Performance verification
To test the performance of the Radical-7 Pulse CO-Oximeter following repairs or during
routine maintenance, follow the procedure outlined in this section. If the Radical-7 Pulse
CO-Oximeter fails any of the described tests, discontinue its use and correct the problem
before returning the unit back to the user.
Before performing the following tests place the Radical-7 Pulse CO-Oximeter Handheld
into the Docking Station, connect the Radical-7 to AC power and fully charge the Radical-7
Pulse CO-Oximeter Handheld battery. Also disconnect any patient cables or pulse oximetry
probes, as well as SatShare, serial or analog output cables from the instrument. Set the
Radical-7 Pulse CO-Oximeter to normal operating mode by selecting the Home Use param-
eter in the General Menu to “No”.
9
service / maintenanceservice / maintenance