4-10 Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual
4
Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual 4-11
4
operation
SETUP MENU LEVEL 1 – ALARM VOLUME
Push the Mode/Enter button to enter menu level 1.
SETTING
Alarm Volume Use Up or Down Arrow Keys to adjust
parameter to desired setting.
Alarm on/off
SETUP MENU LEVEL 2 – ALARM LIMITS
Push the Mode/Enter button again to enter menu level 2.
SETTING
2X
SpO
2
Low Alarm Limit
Use Up or Down Arrow Keys to
adjust parameter to desired setting.
SpO
2
High Alarm Limit
Pulse Rate Low Alarm Limit
Pulse Rate High Alarm Limit
SpCO Low Alarm Limit
SpCO High Alarm Limit
SETUP MENU LEVEL 3 – AVERAGING AND SENSITIVITY
Push the Mode/Enter button again to enter menu level 3.
SETTING
3X
Sensitivity.
Hi = Maximum
Nor = Normal
APO = APOD
Use Up or Down Arrow Keys
to adjust parameter to desired
setting.
Note: These changes affect SpO
2
monitoring only.
Averaging. The signal averaging
time of this device can be set to:
2*, 4*, 8, 10, 12, 14 or 16 seconds
FastSat*
On, Off
*Select “Yes” to activate the FasSat algorithm. The FastSat averaging time is dependant on
the input signal. FastSat is automatically enabled in 2 and 4 second averaging.
operation
SETUP MENU LEVEL 4 - TREND SETTINGS
Push the Mode/Enter button again to enter menu level 4.
To enable trending of patient data, the trend feature must be enabled (set to ON), and the
current date and time must be set. See Section 4, Trend setup and use.
The current date and time can only be set if the Trend is set to “ON”. The date and time
menu selections are not available if Trend is set to “OFF”.
SETTING
4X
Trend ON / OFF
Use Up key to turn trend ON.
Use Down key to turn trend OFF
Set Month
Use Up or Down Arrow Keys to adjust
parameter to desired setting.
Set Day
Set Year
Set Hour
Set Minute
A valid date must be entered. If an invalid date is entered (i.e. February 31), the trend will
not turn on and “tnd off” will be displayed.
Note: The date and time must be set before trending will be enabled. The Rad-57 will
automatically ‘time out’ of the setup menu after 10 seconds with no key presses. If the
Rad-57 should time-out fo the Trend Settings menu, the trend will not be enabled.
Note: Enabling trend (setting Trend to “ON”) will erase all trend information in the Rad-57
SETUP MENU LEVEL 5 - LED BRIGHTNESS AND FACTORY DEFAULTS
Push the Mode/Enter button again to enter menu level 5.
SETTING
5X
LED Display Brightness (4 levels)
Note: All LED indicators are illumi-
nated while adjusting this setting.
Use Up or Down Arrow Keys
to adjust parameter to desired
setting.
Restore Factory Defaults
Pressing a sixth time returns the Rad-57 to patient monitoring display in the
Saturation/Pulse Rate Mode.